Diagram 1 shows the results of the analysis of a standard sample gas using a Zero-internal-volume
valve and a standard valve. As can be seen from the diagram, the peaks obtained using the
standard valve are obtuse. In addition, the roots of each peak are wide and there are parts in which
some ingredients mix. In comparison, the peaks for the Zero-internal-volume valve are acute and
the widths of the roots of each peak are not wide. The acuteness of the peaks shows that the
performance in separation was of high quality. As a result, we can confirm that the
Zero-internal-volume valve improves the performance of a gas sampling control device and the
accuracy in chromatographical analysis.
Details of the Measuring Method
Sample gas was collected by low-temperature condensation at the concentrator (a trap made of stainless steel)
in a gas sampling control device, and then placed in an analyzer after heating and removal from the concentrator.
Ideally the flow path from the concentrator to the analyzer should be zero.
The circuit for fluid control is composed of solenoid valves as follows:
Separation Circuit Chart
Details of the Separation Conditions
Column: HP-INNOWAX (inner diameter: 0.25mm, length: 30m, cross-linked polyethylene glycol stationary phase of 0.25micro meter)
Carrier gas: He (Pressure: 79 kPa)
Column temperature: 45 degrees centigrade
Gas chromatograph (Detector): GC-17A (Method: FID)
Standard sample: Acetone, Ethanol and Methanol (1ng injection)
Notes:
1. The gas sampling control device that was used for the above experiment was developed in a joint
research project with the Nagoya Institute of Technology.
2. The above experiment was carried out in a joint research project with the Nagoya Institute of
Technology.
Reference:
'Chromatography - Mechanism of Separation and Application' Takao Tsuda, 2nd edition, Maruzen,
1995 (published in Japanese)